LOW RANGE TURBIDITY SENSOR
The low-range turbidity sensor guides the parallel light emitted by the light source downwards
nto the water sample in the sensor. The light is scattered by the suspended particles in the water
sample, and the scattered light at 90 degrees to the incident angle is received by the silicon
photovoltaic cell receiver immersed in the water sample. By calculating the relationship between
the 90-degree scattered light and the incident beam, the turbidity value of the water sample
is obtained.
It is widely used in the online monitoring of turbidity in front of and after the filter, finished
water in the water plant, direct drinking water system, and the online monitoring of turbidity
in various industrial production circulating cooling water, filtered water, and reclaimed water
reuse and other links.
FEATURES
● A continuous reading scattering turbidimeter, specifically designed for low-range turbidity detection
● Data is stable and has good reproducibility
● Simple to clean and maintain
● Modbus RTU digital communication, anti-interference, and easy to configure
TECHNICAL PARAMETERS
Range:0.001-100 NTU
Display accuracy:±2% or ±0.015 NTU of the reading,Take the larger value, @ 0.001-40 NTU; ±5% of the degree @40-100 NTU
Resolution:0.01-0.1NTU depending on the different ranges
Repeatability: ±2% F.S
Flow rate:300ml/min≤X≤700ml/min
Operating temperature: 0-45C (no condensation)
Dimensions:310mm210mm410mm (length * width * height)
Main materials: Body:ABS +SUS316L Seal: Butadiene acrylonitrile rubber Cable:PVC
Waterproof grade: IP66/NEMA4
Power supply voltage: Wide voltage range 9-36VDC,recommended 12VDC
Power consumption: 3.5W for low-power LED version sensors, 15W for tungsten lamp
version sensors
Cable length: 3m standard cable, which can be extended.